Оцінка основних похибок вимірювання показника заломлення аналіту приладом на основі явища поверхневого плазмонного резонансу
Loading...
Date
Authors
item.page.orcid
DOI
item.page.thesis.degree.name
item.page.thesis.degree.level
item.page.thesis.degree.discipline
item.page.thesis.degree.department
item.page.thesis.degree.grantor
item.page.thesis.degree.advisor
item.page.thesis.degree.committeeMember
Journal Title
Journal ISSN
Volume Title
Publisher
НТУ "ХПІ"
Abstract
The main sources of measurement errors of the refractive index of the analyte for the device based on the phenomenon of surface plasmon resonance "Plasmon-6" were identified: laser diode wavelength, temperature control operation, surface topology of sensing element metal film. Based on numerical analysis, it was determined that the largest contribution to the measurement error of the refractive index of the analyte component makes temperature error δNT= 54,24 • 10⁻⁵, and the least - the surface roughness of the metal layer sensitive element δNS = 8·10⁻⁵. The error of approximation for the laser wavelength of 650 nm is δNA = 21·10⁻⁵. It defines the main directions of the major decrease measurement errors are: increasing wavelength excitation of surface plasmon resonance, reducing the surface roughness of the metal layer of the sensing element and stabilization of temperature modes of the device. The practical significance is that the results of numerical analysis can be used to improve existing and create new devices based on surface plasmon resonance.
Description
Keywords
Citation
Дорожинський Г. В. Оцінка основних похибок вимірювання показника заломлення аналіту приладом на основі явища поверхневого плазмонного резонансу / Г. В. Дорожинський // Вісник Нац. техн. ун-ту "ХПІ" : зб. наук. пр. Темат. вип. : Механіко-технологічні системи та комплекси. – Харків : НТУ "ХПІ". – 2015. – № 36 (1145). – С. 20-22.